BodyPartExamined correction
Set the correct anatomical region in BodyPartExamined.
Problem
A head CT was incorrectly tagged with Body Part Examined = “CHEST” due to a technologist selection error at the scanner console. This mislabeling causes the study to appear in the wrong hanging protocol, be excluded from body-part-specific queries, and potentially cause mismatched priors to be pulled for comparison.
Steps
-
Open the file (⌘O) and switch to Edit mode (⌘3).
-
Filter for Body Part Examined (0018,0015) and double-click to edit.
-
Enter the correct body part — e.g.
HEAD. The CS VR (Code String) accepts any short code string; standard values include HEAD, CHEST, ABDOMEN, PELVIS, EXTREMITY. -
Click Save change.
-
If you’re also correcting Laterality (0020,0060), filter for it separately and edit — e.g. set to
L(left),R(right), orB(bilateral).
Result
Body Part Examined shows its corrected value in gold, so the PACS hanging protocol selects the right layout. Body Part and Laterality edits are tracked and undoable independently.