Metadata correction

Laterality fix

Correct left/right laterality on a study acquired with the wrong side labeled.

Problem

A knee X-ray or breast tomosynthesis study has Laterality set to blank or wrong, e.g. the left knee was imaged but the tag says “R”. Incorrect laterality can cause the wrong anatomical side to be displayed, lead to wrong-site surgery concerns, and break comparison with prior studies of the same side.

Steps

  1. Open the file (⌘O) and switch to Edit mode (Ctrl+2 / ⌘2 in the desktop app).

  2. Filter for Laterality (0020,0060) and double-click to edit.

  3. Enter the correct value: L (left), R (right), or B (bilateral). The CS VR accepts any short string.

  4. Click Save change.

  5. Also check and correct Image Laterality (0020,0062) if present. This tag provides per-image laterality for paired body parts.

Result

Laterality shows its corrected value in gold. Series-level Laterality (0020,0060) and instance-level Image Laterality (0020,0062) can be corrected independently.